Application profile
|
The
universally applicable process-controlled
hand-held instruments, Series T900, are ideal
for measuring operations in which high
accuracy counts or the possibility of online
documentation is demanded. |
|
Areas of
application in industry &
laboratory for example |
- For
taking measurements for quality assurance under
ISO 9000 - As a reference instrument for
checking production - For taking comparison
measurements in service and repair - For
registration of humidity and temperature in air
conditioning and environmental engineering
- Long-period monitoring of temperature and/or
relative humidity with online
documentation |
|
Product features
|
- RS232
interface galvanically isolated - Optional
SmartGraph Windows Software for graphic and tabular documentation - High accurate
analog output (0-1 Volt) - High measuring
accuracy - Integrated calibration function for
simple compensation of sensor tolerances -
Physical 1-point, 2-point or 3-point calibration
function - Measuring channels are freely
assignable - Recording maximum, minimum, hold
and average values - Integrated sensor holder
for one hand operation - Mains operation
possible - Simultaneous display of two
measured values - Differential temperature
display |
|
T905/T955
|
The two
multi-function-instruments are outstanding for
their high accuracy of ±0,03°C. You can
optional plug in Pt100 or thermocouple (type
J, K, L, N, R, S, T), as well as probes for
humidity and flow. The high precision makes them eminently suitable as reference
instruments for testing subordinate measuring
instruments in quality assurance laboratories
or for checking temperature-critical processes. These instruments are frequently
provided with a DKD calibration
certificate. |
|
Technical data
|
|
|
Inputs
|
2-channel, Pt100, thermocouple:
type K, J, L, N, R, S, T, humidity, dewpoint,
flow |
|
Measuring range |
|
|
Pt100 |
-200°C...+850°C |
|
Thermocouple
|
according
DIN (-200°C...+1760°C) |
|
Ohm
|
0...400
Ohm |
|
Humidity
|
0 %...100
%rH |
|
Flow |
0...40
m/s |
|
Accuracy
|
|
|
Pt100
|
±0,03°C
from -100°C...+150°C ±0,05°C from
-200°C...+200°C otherwise 0,1 % |
|
Thermocouple R, S |
±1,0°C
+0,1 % |
|
Thermocouple K, J, L, N, T
|
±0,3°C
from 0°C...+200°C ±1,0°C to 1000°C ±1,5°C
remaining range |
|
Humidity
|
±1,5
%rH |
|
Flow |
0,5 % of
end value |
|
Ohm
|
0,5 % of
measuring range |
|
Resolution |
0,01°C
from -200°C...+200°C otherwise 0,1°C
respectively 0,1 % 0,01 m/s |
|
Connectors |
DIN
8-pole |
|
Working temperature |
0°C...+40°C |
|
Display |
2-line
LCD |
|
Housing |
plastic |
|
Dimensions |
200 x 200
x 80 mm (LxBxH) |
|
Weight |
950
g |
|
Power supply |
230
VAC |
|
Order No. |
5000-0955
(T905: 5000-0905) |
Downloads
T900
Spec |